12L Portable Suction Machine (Medical Aspirator)

->15L Portable Suction Machine DSU-9EAItem: DSU-DX12
Detail:

Fashion Style Aspirator Suction

 

Model

● DSU-DX12   Aspirator Suction        12L/min   1000ml Suction bottle

  

Functions

The DX Series Aspirator is a very handy small suction unit. It is driven by an electromotive, maintenance-free diaphragm-type pump. During operation the pump generates a vacuum within the hose system and the collection jar. Thus sucking off secretions or fluids (e.g. By means of a suction catheter). The fluid is gathered in the collection jar. A mechanical overflow safety (on the inner part of the collection jai lid) avoids penetration of secretion into the pump head. The final vacuum and following, the air-flow rate can be adjusted by means of the fine control and the vacuum-gauge.

 Features

>>The unit can be equipped with a rechargeable battery (accumulator). This battery is optional.

>>Integrated microprocessor-based technology assures safe charging of the battery; overcharging is thus impossible.

      >>An over temperature stop controlled by electronics avoids overheating of the unit.

>>A disposable bacterial filter plate integrated in the lid of the collection jar prevents bacteria and liquid from penetrating into the pump.

 

Accessories

In addition to the basic device, the scope of delivery comprises the following parts:

-Silicone suction hose Ø6mm, L=1.30m 

-Hose connector (fingertip)

-Mains cable for 230V /50-60Hz

      S M L size suction catheters

 

Packing information

Color gift box

Unit gross weight: 3.5kg

2 boxes /Carton, 9kgs 

Carton Size: 46.5*40*34cm

     

Technical specifications

 

Model

DX-12


 

 

Air flow rate pump

121/min


 

Max. vacuum

75kPa+5kPa


 

Vacuum readout

-1…0 bar (±25mbar)(mm Hg; kPa)*

Additional air regulation

Mechanical regulating valve

Collection jar

1I secretion container,1I Receptal ®container system or

1I MediVac container system

Suction hose

Ø 6mm,1.30m length

Voltage

230V~(+/-10%)50Hz

Current input(max)

Ca. 1.25A(230V~)

Power consumption

Ca. 300 VA(230V~)

Power cable

2 m

Operating time

Interrupted use over approx. According to the environment 30-45 min at 230V~,50Hz,20℃;Cooling period approx. 60min,depending on ambient temperature

Fuse

T 2. 5A/H(230V~);

Protective earth conductor resistance

Ear leakage current

Enclosure leakage current

Patient leakage current

-

-

N.C. less than 0.1mA

-

Heat emission

100J/s

Noise level

50.0dB(A)@1m(nach ISO 7779)

Ambient conditions

Transport/storage

 

 

Operation

-30…+50

5...90%humidity

non-condensing air pressure 700...1060 hPa

+10…+35

20...80% humidity,

non-condensing air pressure 700...1060 hPa

Weight

3.5kg


 

Regular safety relevant inspections

Recommended: once yearly

Protection class(EN) 60601-1

II

Degree of protection

Typ BF

Protection category

IPX 0

Classification acc. to Annex IXEEC

Directions 93/42/EEC

IIa

CE marking

CE 0123

Rules applied

EN 60601-1-1:1990+A1:1993+A11:1993+A12:1993+A2:1995+A13:1996

EN 60601-1-2:2006

EN 61000-3-2:2006

EN61000-3-3:1995+A1:2001+A2:2005

EN ISO 10079-1:1999

UMDNS-Code

10-208(tracheal suction device)

 

*1 bar≈750.06 mm Hg≈1000hPa/depends on daily atmospheric pressure issue of Technical Specifications

 

 